SEM-XRF : Integrated e⁻-Beam / X-ray XRF Elemental Analysis

X-ray Sources for Electron Microscopy (SEM)

microXRF X-ray sources for Scanning Electron Microscopes (SEM): addition of an polycapillary X-ray tube and Iridium Ultra software will transform your SEM’s quantitative analytical capabilities. Higher peak to background ratios enable greater elemental sensitivity for higher Z elements: sensitivity exceeding e-beam excitation by a factor of 10-1000X. Exceptional beam stability, together with a modern SDD X-ray detector, afford higher precision with ppm-level sensitivity. Non-conductive materials may be analyzed without any special preparation or coating. We integrate with your SEM to deliver full spectrum analysis using excitation from both the e-beam and our X-ray source.

download a brochure

What is SEM-XRF?

IXRF’s Xb SEM micro-spot X-ray source adds the capabilities of a complete micro X-ray fluorescence (microXRF) spectrometer to a scanning electron microscope (SEM). Xb SEM fits on an unused or open high-angle port of almost any SEM. Total non-destructive, μXRF users benefit from both superior trace element sensitivity and broader elemental coverage (Na through U using K- and L-lines) .

SEM/EDS vs SEM-XRF

Relative sensitivities of K-lines at 30 kV excitation.

Why SEM-XRF ?

At left is shown a plot of a theoretical silicon sample containing trace amounts of Na-Sn. At 0-2 keV, e-beam (SEM/EDS) is clearly more efficient. Above about 2 keV, X-ray photon excitation (SEM-XRF) becomes orders of magnitude more efficient. This allows for quantification of more elements at lower concentrations than is possible with SEM/EDS.

Advanced microXRF System for SEM

Distribution analysis stores complete spectra for every map point for on- and off-line analysis. Samples can be analyzed with micro-spot X-ray beam and e⁻-beam simultaneously without position change. Both excitation methods are integrated in our analytical software suite. No interference with normal SEM operation, the X-ray source can stay in its measurement position permanently. microXRF is completely non-destructive and does not require the sample to be coated. Use your existing EDS detector system; we support most all microscopes.

Elemental Analysis with a Mapping microXRF Spectrometer

• Analytical results compare to those of standalone μXRF systems
• State-of-the-art analytical software: Iridium Ultra
    – Decades of continuous development & innovation
• Selectable primary X-ray filters
    – To suppress Bremßtrahlung and diffraction peaks
    – For lower detection limits, down to PPM levels for most elements
• Slewing precision XY-stage to create image maps  (image right)
• Allows sample tilt to produce minimum spot sizes
• No special cooling is required, our sources are air cooled
• Samples do not charge up, so no coatings are needed
• Available for most column-type SEMs
• Solves those difficult overlapping peak issues common to SEM/EDS

Slew scan XY-stage

Slew scanning XY-stage

Chemical Composition by Rayleigh to Compton Ratio Method

Rayleigh to ComptonAnother advantage of having micro-XRF in a scanning electron microscope (SEM) is the ability to employ the characteristic lines from the anode material (typically Ag, Cu, Mo, Rh or W) of a X-ray tube to non-destructivity determine the average atomic number of any phase or artifact. The high specificity of elastic (Rayleigh), as well as inelastic (Compton), X-ray scattering to the mean atomic number of a specimen has long been recognized.1 Micro-spot X-ray fluorescence (micro-XRF) may thus be exploited to non-destructively gain more information on chemical composition. In practice, the Rayleigh/Compton intensity ratio from XRF spectra may be calibrated, with standard reference materials, relative to mean atomic number. This sort of calibration curve can reveal valuable information on the elemental composition complementary to that obtained from conventional elemental analysis via XRF-SEM and/or SEM/EDS. Particularly for matrices of lower mean atomic numbers (like plastics), the sensitivity of the approach is so high that it can be easily distinguish between specimens of mean atomic numbers differing from each other by 0.1 Z. Hence, the content of light elements which are “invisible” for XRF, particularly hydrogen, or of heavier impurities/additives in light materials can be calculated “by difference” from the scattering calibration curve.

Xb Product Features

IXRF’s Xb custom micro-spot X-ray source is designed exclusively for use on electron microscopes.  A compact design with slide mounting allow very close coupling to the sample. Patented polycapillary optics focus X-ray  photons down to sample spot size as small as 20µm. Xb is offered in 20µm, 25µm and 40µm spot sizes. An integrated high-voltage power supply operates up to a maximum power of 50 watts (35-50 kV and 1.0 mA depending on anode material). Close coupling provides XRF analytical results comparable to those from traditional “’benchtop EDXRF” or “mainframe” XRF instruments.  An additional benefit is that the Xb X-ray source is designed so that it does not interfere with the normal operation of the electron microscope, including the use of the electron beam on the same sample at the same time so as to simultaneously collect data on all elements.

SEM-XRF Advantage: Quantify Trace Elements

Electron beams (from scanning electron microscopes) produce very high backgrounds, hiding trace elements in a sample. Direct X-ray excitation from a true “X-ray source” don’t have this effect. Using IXRF’s Xb, low ppm levels of elements can be easily identified, quantified … even producing trace level X-ray maps to view elemental distribution of trace elements in your sample. For elements above Na, limits of detection (LOD) can be as low as 10 ppm, depending on element and matrix.

SEM-XRF Advantage: Broad Elemental Coverage

X-ray excitation up to 50 kV, together with a variety of available primary filters, allows for high-sensitivity analysis of first and second row transition metals (using K-lines). Even heavier elements may be similarly analyzed using their L-lines.

Polycapillary Optics' Advantage

Polycapillary focusing optics collect a large solid angle of X-rays from an X-ray source and focus them to a spot as small as 10 µm. The X-ray flux density obtained is a few orders of magnitude higher than that obtained with a conventional pinhole collimator. The main application of these optics is micro X-ray fluorescence (µXRF) analysis, which has been widely used for thin film and plating analysis, precious metal evaluation, alloy measurement, and monitoring of electric circuit board coatings. Use of polycapillary focusing optics will significantly enhance the detection sensitivity and allow the high performance to be achieved with low-powered X-ray tubes. Micron-sized spatial resolution makes small feature evaluation possible for electronics and precious metal applications. Polycapillary optics offer 100x-10,000x gain with output focal spot sizes as small as 10 µm.

Product Specifications

Xb

ANODE TYPE

Side-window

TARGET MATERIAL

Ag, Cu, Mo, Rh & W

ACCELERATING VOLTAGE

0-50 kV

BEAM CURRENT

Max 1 mA

EXCITATION SPOT SIZE

20, 25, 40 µm

COLLIMATOR SIZE

Patented polycapillary focusing optic

SOURCE FILTERS

Available upon request

COOLING REQUIREMENTS

Air cooled (fan)

CONTROLS / SAFETY

Variable control kV/µA, X-ray on/off buttons, kV/µA display, internal interlocked shutter. Interlocked to SEM, keyed power-on switch, HV-On lamp, warning beacon

ƒX Product Features

IXRF’s new ƒX SEM™  is an affordable, air-cooled custom X-ray source option designed exclusively for use on electron microscopes.  With a compact design and slide mounting, very close coupling to the sample is achieved. This unique design delivers “high flux” X-rays to small or large excitation areas on a sample’s surface. The ƒX SEM™ offers excitation spots ranging from 500 µm to 25 mm (ellipse minor side width).

Supported On All SEM Brands

The integrated high-voltage power supply operates up to a maximum power of 10 W (35 kV and 0.1 mA depending on anode material). Close coupling provides X-ray fluorescence (XRF) spectrometry analytical results comparable to those from a traditional benchtop EDXRF operating under a vacuum atmosphere. The ƒX SEM™ is designed so that it does not interfere with the normal operation of the electron microscope, including the use of the electron beam on the same sample, at the same time collecting all elements simultaneously.

Non-destructively Quantify and Map Trace Elements

Electron beams (from scanning electron microscopes) produce very high backgrounds hiding the trace elements in the sample. X-rays, from a true “X-ray source” don’t have this effect. Using the ƒX SEM™ low ppm levels of elements can be easily identified and quantified … even producing trace level X-ray maps to view elemental distribution of trace elements in your sample.


Product Specifications

ƒX

ANODE TYPE

End-window transmission

TARGET MATERIAL

Ag, Mo & W

ACCELERATING VOLTAGE

10-35 kV

BEAM CURRENT

0-100 µA

COLLIMATOR SIZE

200, 500 and 1000 µm (others available)

SOURCE FILTERS

Available upon request

COOLING REQUIREMENTS

Conduction cooled, no fan required

CONTROLS/SAFETY

Variable control kV/µA, X-ray on/off buttons, kV/µA display. Interlocked to SEM, keyed power-on switch, Integrated high-voltage power supply, HV-On lamp, warning beacon

X-ray Source + SEM =

  • Sensitivity: Higher peak to background ratios enables greater elemental sensitivity for higher Z elements; sensitivity exceeding SEM/EDS by a factor of 10-1000x.
  • Precision: The extreme X-ray beam stability of Xb, coupled with premium SDD X-ray detectors, yields highest possible precision and lowest achievable detection limits.
  • Accuracy: Using elemental K-lines (and L-lines of high-Z elements) means less complex deconvolution of various peak overlaps.
  • Non-destructive: Analysis of non-conductive materials without coating or e⁻-beam damage.
  • Full elemental coverage: Integrated with EDS software for the most accurate low energy-high energy (full spectrum) analysis available (by combining e⁻-beam and X-ray beam excitation).

Fits Your Microscope

X-ray tube additions may be interfaced to almost any brand and model of SEM. High angle (35º) as well as horizontal flange mounts are available for every tube package.

Polycapillary Optics

IXRF’s Xb delivers the addition of focusing capillary optics to enable microspot XRF analysis down to 20 microns. A greatly reduced spot size increases micro-XRF X-ray elemental mapping capabilities and enhances trace elemental and/or thin-film analysis at the micron level.

focusing polycapillary optic

20 μm Spatial Resolution

This is a picture of a 12 micron Ni grid X-ray map. The grid is imaged with a 20 micron Xb X-ray source, demonstrating exceptional spatial resolution.

Combined EDS and XRF X-ray mapping bring previously unseen capability to the electron microscopist.

X-ray Tube Automation

Packaged tubes may be operated up to 50kV and come with integral high voltage power supplies. Power supplies are integrated with interlocking vacuum sensors which shut off the x-rays when the SEM chamber is vented. Each unit automatically “ramps” the tubes for increased longevity.

Miniaturization

The chief component of low cost X-ray Fluorescence brought into the electron microscope chamber is the ƒX SEM™ low powered transmission target X-ray tube. The tubes are reduced in size and produce less heat than larger, higher powered tubes. Mo, W and Ag thin film targets deposited on a Be window are available for optimizing varying applications.

X-ray source & Electron Beam Integrated Acquisition

Combined Analysis

Electron Beam Excitation

In order to acquire the most comprehensive and accurate full spectrum in seconds, use your EDS system to capture the light elements between 0-3kV; use the iXRF SEM-XRF tool to acquire the heavier, higher energy elements between 3-50kV.

Combined Quantification

In order to acquire the most comprehensive and accurate full spectrum in seconds, use your EDS system to capture the light elements between 0-3kV; use the iXRF SEM-XRF tool to acquire the heavier, higher energy elements between 3-50kV.

X-ray Source Excitation

Acquire your full spectrum using one technique at a time or run in simultaneous mode. Acquire light and heavy elements using two techniques at the same sample location without ever opening the chamber.

Available Methods

  • Auto
  • ZAF
  • FP
  • Least Squares
  • Match

SEM-XRF Dual Excitation

The IXRF SEM-XRF option, offers the ability to use both the electron beam and Micro-XRF beam simultaneously. With the electron beam at 3-5kV, only the light elements are excited. There is no significant background added to the XRF spectra, allowing all the elements in the sample to be seen.

Applications

XRF provides non-destructive analysis of various sample sizes and sample types including solids, environmental samples, powders and residues. Below are examples of common applications for traditional tabletop (bench top) XRF that are now available inside the SEM:

  • Art and Archeological
  • Chemical
  • Coatings and Thin Film
  • Cosmetic
  • Educational
  • Environmental
  • Food Applications
  • Forensics
  • Metal and Ore
  • Mineral and Mineral Products
  • Petroleum EDXRF
  • Pharmaceutical Applications
  • Plastics, Polymers, and Rubber
  • Plating and Plating Baths
  • Wood Treating
  • Concrete Treating
  • And Others
integrated XRF
SEM-EDS-XRF

Mapping

Trace elemental mapping is possible for nearly all of the above applications. XRF mapping can be performed using either of our products; shown here by this mineral map using the X-Beam.

Sensitivity: Glass

XRF can be used for trace spectral acquisition down to low PPM levels. XRF can be 10-1000 times more sensitive than SEM-EDS analysis, exemplified here by the greatly increased trace level peaks in the NIST SRM 610 glass standard.

Sensitivity: Steel

XRF can be used for trace spectral acquisition down to low PPM levels. XRF can be 10-1000 times more sensitive than SEM-EDS analysis, exemplified here by the greatly increased trace level peaks in a Brammer B.S.35D steel standard.

Dual Power

These are example maps of both X-ray and electron beam excitation are employed simultaneously on the same sample at the same time! Never before has this ability been demonstrated.

Reference Documents

Search Terms

    • Micro‐XRF excitation in an SEM
    • Micro-Focus X-Ray Fluorescence (μ-XRF)
    • Micro-XRF-SEM
    • Micro‐focus X‐ray source for XRF analysis with the SEM
    • Micro‐focus XRF at the SEM
    • Micro-XRF in Scanning Electron Microscopes
    • µ-XRF at a SEM
    • µXRF SEM
    • MicroEDXRF SEM
    • µEDXRF SEM
    • Micro-XRF SEM
    • Micro-EDXRF SEM
    • Polycapillary SEM
    • Poly-capillary SEM
    • SEM/XRF
    • XRF/SEM
    • SEM-XRF
    • XRF-SEM
    • XRF SEM
    • X-ray source SEM
    • SEM X-ray source
    • X-ray SEM
    • X-ray tube SEM
    • SEM X-ray
    • SEM X-ray tube
    • SEM polycap
    • polycap SEM
    • SEM polycap
Interested in what we do?

CONTACT US

  • Hi, my name is
  • and I am interested
  • in the
  • Please contact me at
  • or call me at
  • This field is for validation purposes and should be left unchanged.